Drop Counter Calibration
Use the procedure below to calibrate drops so that a precise volume of titrant is recorded in units of milliliters.
Calibrating Your Drop Counter
- After the Drop Counter has been set up, choose Calibrate > DIG 1: Drop Counter from the Experiment menu.
- Select Automatic.
- Place
a 10 mL graduated cylinder directly below the slot on the Drop Counter,
lining it up with the tip of the reagent reservoir.
- Open the bottom valve on the reagent reservoir (vertical). Keep the top valve closed (horizontal).
- Slowly
open the top valve of the reagent reservoir so that drops are released
at a slow rate (~1 drop every two seconds). You should see the drops
being counted on the computer screen.
- When the volume of titrant in the graduated cylinder is between 9 and 10 mL, close the bottom valve of the reagent reservoir.
- Enter
the precise volume of titrant (read to the nearest 0.1 mL) in the edit
box. You may want to record the number of Drops/mL displayed on the
screen for possible future use.
- Click OK. Discard the titrant in the graduated cylinder as indicated by your instructor and set the graduated cylinder aside.